home > about > radiography (local rules)
radiography These
local rules are effective 27.7.2006; amended 10.4.2008
local rules - web version 2.0
Every dental practice with radiographic (x-ray) equipment is required to provide a set
of “local rules”. These record all the working practices we must follow to ensure that we are safe
when working with radiation and that we comply with the various regulations governing radiation in
dentistry. This document contains the “local rules” that apply to hygeia dental
care’s premises at Malt Mill Lane, Totnes.
duties of employees
All clinical personnel must:
a) Ensure that exposures to staff, and all other persons, are kept as low as
reasonably practicable;
b) Take reasonable care when working with any aspect of dental radiography -
in particular, radiography must be conducted with due regard to minimising
accidental or unintended doses to patients;
c) Immediately report to the RPS (Joanne Giddy) any incident that may result
in the overexposure of themselves or anyone else.
All clinical personnel have a duty to exercise reasonable care and act responsibly in relation to
radiography. They are required to know and observe these rules at all times.
x-ray equipment at Malt Mill Lane
There are two x-ray units at Malt Mill Lane:
unit 1
| location |
treatment room 1 |
| manufacturer |
Trophy Trex |
| model |
IRIX 70 – Trophy CCX |
| serial no. |
875097 |
| manufactured |
23/6/98 |
| installed |
15/10/98 |
| operating potential |
71kV |
| total beam filtration |
2.9mm Al |
| beam profile |
32mm x 42mm |
| focal spot to skin distance |
200mm |
| optimum dose for imaging |
0.70mGy (digital sensor - adult mandibular molar) |
| patient skin dose |
1.10mGy (digital sensor - film type 6 - adult mandibular molar) |
unit 2
| location |
treatment room 2 |
| manufacturer |
Trophy Trex |
| model |
IRIX 70 – Trophy CCX |
| serial no. |
921048 |
| manufactured |
20/7/99 |
| installed |
8/99 |
| operating potential |
69kV |
| total beam filtration |
2.9mm Al |
| beam profile |
54mm diameter |
| focal spot to skin distance |
200mm |
| optimum dose for imaging |
1.20mGy (film speed E - adult mandibular molar) |
| patient skin dose |
1.20mGy (film speed E - film type 4 - adult mandibular molar |
important note: The garden area outside must be unoccupied
whilst x-rays are being taken
It is the responsibility of Joanne Giddy (the RPS – see below) to notify the Health and Safety Executive of
the use of, and any changes to, the radiography equipment in use at the premises.
radiation protection adviser & medical physics expert
The Health Protection Agency Radiation Protection Division (formerly known as the NRPB)
has been appointed as Radiation Protection Adviser (RPA) and Medical Physics Expert (MPE) to hygeia dental
care in respect of the premises at Malt Mill Lane, Totnes in accordance with Regulation 13 of the Ionising
Radiations Regulations 1999 (the “IRR”) and the Ionising Radiation (Medical Exposure) Regulations 2000
("IRMER"). They can be contacted on 0113 269 9643.
radiation protection supervisor
Joanne Giddy is the appointed Radiation Protection Supervisor (RPS) for hygeia dental care’s
premises at Malt Mill Lane, Totnes. It is her responsibility to ensure that all persons working at the
practice observe these local rules and that they comply with the IRR, any other relevant statutory instruments
and relevant guidance published by competent professional and statutory bodies.
The RPS (Joanne Giddy) must undergo training in radiography at least once
every five years to maintain an
appropriate “Core of Knowledge” as specified in the Schedule to the Ionising Radiation (Protection of Persons
Undergoing Medical Examination or Treatment) Regulations 1988 (“POPUMET”).
It is the responsibility of the RPS to periodically review and (in consultation with the RPA/MPE)
update these local rules.
controlled and supervised areas
The IRR specify “controlled” and “supervised” areas in relation to dental intra-oral x-ray
equipment.
controlled areas
When either x-ray unit is operating, a “controlled area” exists. A controlled are is somewhere that you MUST NEVER GO
while the x-ray unit is in operation. These controlled areas are:
a) within the primary beam until it is sufficiently attenuated by distance
or absorption in material
The “primary beam” is the “beam” of x-ray emissions (ie radiation) that emerges from the head of the
x-ray unit (the large part on the end of the arm). The beam is directional, like a car headlight,
and travels in the direction that the x-ray head is pointing.
To avoid being “within the primary beam” you must never stand anywhere in the room where the x-ray head
is pointing at you or even vaguely in your direction - just as you are not in the beam of a car’s headlights
if you are standing behind the car.
Always stand where the operator (the dentist) tells you. This will usually be at the far end of the
surgery next to the door (for more on the correct operator’s position, see “use of x-ray equipment”, below).
It is up to the operator to ensure that no matter what type of x-ray is being taken, you are outside
the primary beam.
Of course, as the regulations state, the controlled area ends once the beam has been “sufficiently
attenuated” – ie reduced to a harmless level. The beam does not go on forever - even the molecules in
the air will stop radiation traveling more than a certain distance. X-rays are stopped even more
quickly when they pass through solid materials like concrete and especially lead.
The wall between the two surgeries is lead lined. Accordingly, if you are standing in surgery 1 and
someone in surgery 2 points the x-ray unit directly at you through the wall only a few centimetres away,
the radiation will not do you any harm because the lead absorbs it. This means that you don’t have to
worry about what the x-ray unit in the other surgery is doing.
b) in all directions within 1.5 metres of the x-ray tube head and the patient
Even if you are effectively “behind” the x-ray head (ie it is pointing away from you) you still need to
stand a minimum distance away from it. While 1.5 metres is the recommended safe distance for x-ray equipment operating at 70kV (as ours does) the
practice adopts a distance of 2 metres. Standing near the surgery door at the far end of the room places you at a sufficient distance from the
x-ray source, but always follow any instructions from the operator as to exactly where you should stand.
supervised areas
There are no supervised areas outside the controlled areas defined above. All that
clinical personnel need to know about supervised areas is that there are none. Just be sure to stay
out of the controlled areas.
classified persons and written system of work
There are no “classified persons” on our team and the provisions on written systems of work
do not apply.
x-ray equipment – testing, servicing, maintenance & repair
All x-ray units at Malt Mill Lane must be subjected to a radiation safety assessment
at least once every 3 years. It is the responsibility of the RPS (Joanne
Giddy) to ensure that this is done. Radiation safety assessments are
carried out using standard HPA test packs. It is the responsibility of the RPS (Joanne Giddy) to ensure that any recommendations made as a result of
the assessments are acted upon within 3 months.
All x-ray units at Malt Mill Lane must undergo routine servicing at least
once every year. It is the responsibility of the RPS (Joanne Giddy) to
ensure that this is done. Servicing is undertaken by Blackdown Dental
Equipment Services who can be contacted on 01823 672241 or 07834 549560.
All x-ray units at Malt Mill Lane must undergo routine surveillance at least
once every 6 months. Routine surveillance checks the correct functioning of the audible and visual
warning systems, exposure controls, x-ray arm counterbalance mechanism and condition of the x-ray tube
head. It is the responsibility of the RPS (Joanne Giddy) to
ensure that this is done and to arrange for any necessary repairs to be carried
out. Repairs are undertaken by Blackdown Dental
Equipment Services who can be contacted on 01823 672241 or 07834 549560.
No-one is permitted to interfere with or modify any part of the x-ray equipment unless they have first
referred the matter to the RPS (Joanne Giddy).
Following any testing, servicing, maintenance or repair work, no x-ray unit
or ancillary equipment may be accepted back into service until the RPS (Joanne
Giddy) has reviewed the service report. This should confirm that the
equipment has been left in a state fit for use and that no alterations have been
made which may significantly affect patient doses. If such alterations
have been made, the RPS (Joanne Giddy) should seek advice from the RPA/MPE
before bringing the equipment back into use, since a critical examination may be
required first.
use of x-ray equipment
The x-ray equipment must only be operated by a suitably qualified person, or by a person
under their supervision who has received adequate training and instruction in the use of the x-ray
equipment. At Malt Mill Lane only Joanne Giddy is qualified to operate the equipment.
The exposures used for radiography must be no greater than those required to yield correct radiographic
densities with complete development of the film. The exposure timings for both x-ray units are
pre-set. The correct film speed settings are selected using the “f” button on the control box.
The practice uses a digital sensors. The correct setting for these is film type
4 in surgery 1 and film type 3 in surgery 2. The controls
must also be set for the appropriate anatomical view. For other film types and guidance on adjustments
to exposure times to compensate for patient size, refer to the IRIX 70 user’s manual.
The operator of the x-ray equipment must take up a position outside the main x-ray beam and where the
instantaneous dose rate does not exceed 7.5 microsieverts per hour. For both x-ray sets the appropriate
operator position is 2 metres away from and behind (ie in the rear 180 degrees of) the x-ray tube.
If the operator stands inside the surgery next to the door and as near to the end wall of the surgery as
possible, this should ensure that the operator is in a safe position provided that the patient and x-ray
tube have been correctly positioned. However, it is the responsibility of the operator to check this
and ensure that they, their DSA and any other persons present in the room are located at least 2 metres away
from, and in the rear 180 degrees of, the x-ray tube head.
The operator must also ensure that the x-ray tube head is never directed toward a
window, a door or an unshielded partition wall.
If this appears to be the case, the patient must be repositioned before the exposure is taken so that the
tube head is pointing toward an external wall or shielded internal wall.
The operator must ensure that any personnel who do not routinely work in the surgery (reception and
administration personnel, for example) are not present in the room when x-rays are taken.
The operator is supplied with and is required to use x-ray film holders with integral beam alignment
devices. These have been proven to reduce the number of radiographs that are judged unacceptable by
reason of positioning errors, thereby reducing the overall number of x-rays taken and x-ray exposure for
clinical personnel and patients alike.
During every radiographic examination the operator must observe the radiographic warning light to check
that the exposure terminates correctly. X-ray sets must be turned off (using the switch on the control
panel) when not in use.
damage to x-ray equipment
If an x-ray unit is damaged, immediately isolate the unit from
the mains electricity supply. To do this, simply switch the power off
using the switch on the front panel of the x-ray control unit. The x-ray
control unit is the box on the wall near the surgery entrance door (to which the
remote switch is attached). Next, switch off the supply circuit labeled
"X-ray unit" (circuit breaker number 3) at the consumer unit in the relevant
surgery. Notify the RPS (Joanne Giddy) of the damage immediately. Do
not use the x-ray unit again until the malfunction has been investigated by the
RPS (Joanne Giddy)
and any necessary remedial action has been taken by service agents. Servicing is undertaken by Blackdown Dental
Equipment Services who can be contacted on 01823 672241 or 07834 549560.
contingency plans
If the exposure warning light remains on after the set time has elapsed, or any other
fault is indicated or suspected, immediately isolate the x-ray unit from the mains electricity supply.
To do this, simply switch the power off using the switch on the front panel of the x-ray control unit.
The x-ray control unit is the box on the wall near the surgery entrance door (to which the remote switch is
attached). Next, switch off the supply circuit labeled "X-ray unit"
(circuit breaker number 3) at the consumer unit in the relevant surgery. Notify the RPS (Joanne Giddy) of the malfunction immediately. Do not use the x-ray unit again until
the malfunction has been investigated by the RPS (Joanne Giddy) in consultation
with the RPA/MPE, a satisfactory explanation has been found
and remedial action has been taken by service agents. Servicing is undertaken by Blackdown Dental
Equipment Services who can be contacted on 01823 672241 or 07834 549560.
accidental over exposure
If an x-ray unit malfunctions or appears to be defective and it is suspected that a patient
undergoing a medical exposure may have been exposed to ionising radiation to an extent much greater than
intended, the incident must be reported immediately to the RPS (Joanne Giddy). She must investigate
the incident immediately in consultation with the RPA (ie the HPA). Unless the investigation shows beyond reasonable doubt that no such incident occurred, it must be
notified to the Health and Safety Executive on 0117 988 6000. Arrangements must be made for a detailed
investigation of the circumstances of the exposure and an assessment of the dose received. A copy of
the report of any investigation must be kept for at least 50 years from the date on which it was made.
radiation monitoring (personal dosimetry)
The practice requires all clinical personnel to wear personal radiation dosemeters when
at work.
The x-ray equipment on the premises is regularly checked to ensure that it is performing in accordance
with HPA (Health Protection Agency - Radiation Protection Division) guidelines. Lead shielding is also
built into the walls of the building, where necessary. This should ensure that nobody is exposed to
more than a minimal level of radiation.
However, x-ray units such as the ones at the practice only need to be checked once every three
years. We felt it wise to introduce personal dosemeters, which are checked every three months,
so that everyone could be confident that they were safe from excessive levels of radiation in their working
environment. To date, none of the radiation dosemeters has produced any significant reading at all.
In other words, no-one working at the practice over the last 9 years has received any more than normal
background radiation.
The practice is obliged to provide personal dosemeters to individual clinical personnel who regularly take
more than 100 intra-oral x-rays each week. No individual at the practice regularly takes that many
x-rays (or is present while they are taken) in a single week. Nonetheless, we consider the use of
personal dosemeters to be an important part of our health and safety regime and so all clinical personnel
are required to wear them.
To ensure that the dosemeters work properly and record useful data, all clinical team members must
follow the procedures set out below:
a) You must wear your dosemeter at all times when at work.
b) Dosemeters must not be taken outside the building (nb: it is acceptable to wear them outside for very
short periods such as when you go to the wheelie bins, but you must leave them behind if you go out for
lunch, for example).
c) When you leave work, your dosemeter must be left at the practice – even if you are just going out for
lunch or to an appointment. Put it on again when you return.
d) If you accidentally take your dosemeter away from work, you must record that fact in the “DOSEMETER
INCIDENT RECORD” – a thin blue book that is kept in the reception desk. Remember to write down your
name, the date and times and a description of where you went.
e) You must also enter any other relevant incidents in the “DOSEMETER INCIDENT RECORD” – e.g.: if you drop
your dosemeter into a chemical solution or leave it in the surgery when you are on holiday (this is
important because the dosemeter may then be repeatedly exposed to higher radiation doses than you would
normally receive and give an incorrect reading).
f) Dosemeters are very expensive to replace – take care not to lose or damage them.
g) Wear your dosemeter at waist level (on the trunk of your body).
h) You can attach your dosemeter to your uniform using the plastic strap and clip supplied or you may
instead use a safety pin.
It is the responsibility of the RPS (Joanne Giddy) to review the results of the monitoring system.
If it appears that doses have significantly increased in relation to any member of the team, the RPS
(Joanne Giddy) must consult the RPA and consider whether that team member needs to be designated as a
classified person for the purposes of the regulations. The RPS (Joanne Giddy) must also consider
whether any changes to working practices are required.
dose investigation level
The formal dose investigation level set by this practice is 1mSv
(refer to
Radiography Risk Assessment - see section on IRR, regulation 21: Dose Assessment).
Should any member of staff receive an annual dose exceeding 1mSv, the RPS
(Joanne Giddy) must conduct a formal investigation in liaison with the RPA/MPE
in order to establish why. The results of any such investigation must be
kept for at least 2 years.
pregnancy - clinical personnel
All clinical personnel must inform the RPS (Joanne Giddy) as soon as
they become aware that they are pregnant. This is because working with ionising radiation carries
certain risks for a developing fetus. Compliance with these local rules
should normally mean that no special precautions are necessary. However,
the RPS (Joanne Giddy) will consult the RPA/MPE
and, if required, implement changes to working practices for the pregnant team member
to ensure that the dose to the fetus of any member of staff is unlikely to
exceed 1mSv during the declared term of pregnancy.
pregnancy - patients
Pregnant patients only present safety issues in connection with dental
radiography when a radiographic examination would result in irradiation of the
pelvic area by the primary x-ray beam (eg where a vertex occlusal projection is
required). We do not use such projections on adult patients and therefore
this can never be an issue.
Nonetheless, we always identify pregnant patients as part of our
clinical risk management procedures by means of medical history taking
(treatment code EMED1) and medical history updates (treatment code
EMED2).
Where a patient knows or suspects that they are pregnant, then they
are - unless there is an urgent need for treatment and radiographic
examination is essential - advised that they may defer radiographic
examination if they prefer.
assistance with radiography
Clinical personnel must never support the x-ray tube head, or
the patient, or hold a film in position for the patient. Whenever a
patient requires assistance during a radiographic examination seek the advice of
our Radiation Protection Adviser – this is the HPA who can be reached on
0113 269 9643.
film processing
The practice no longer uses wet film systems, so all sections in older editions of these
local rules relating to film processing, preparation of solutions, storage of solutions, replacement of
solutions, processing times and temperatures and mounting of radiographs no longer apply.
radiographic quality control
When reviewing x-rays, the dentists must note any deterioration in image quality. If
there is any deterioration in quality, they must consider whether this is a result of errors in positioning
or exposure.
Errors of positioning should be virtually eliminated by the beam alignment devices provided. If not,
the operator must adjust his or her technique to improve results.
Exposure times are pre-set on the x-ray units at the practice. This minimises the possibility
of error.
Processing errors are no longer possible (because we use a digital system),
so a loss of image quality may indicate a malfunction in the x-ray unit and the
RPS (Joanne Giddy) must be informed. It is her responsibility to
investigate the matter and, if necessary, arrange for the x-ray unit, sensors
and software to be checked if she suspects there may be a fault. In her
absence, the dentist should consult the RPA. If it is necessary to arrange
for the unit to be checked, contact Blackdown Dental Equipment Services ltd on
01823 672241 or 07834 549560.
The practice aims to achieve the following x-ray quality targets:
a) Not less than 70% of x-rays should be of "excellent quality" - with no errors of exposure or
positioning
b) Not more than 20% of x-rays should be of only "acceptable quality" - with some errors of exposure or
positioning but which do not detract from the diagnostic utility of the radiograph
c) Not more than 10% of x-rays should be of "unacceptable quality" - with errors of exposure or positioning
which render the radiograph diagnostically unacceptable
If an operator becomes aware that the quality of films has fallen significantly below these targets,
they must inform the RPS (Joanne Giddy) immediately. She must then take action to ensure that overall
quality is improved.
The RPS (Joanne Giddy) is responsible for carrying out audits of radiographic
image quality at six monthly intervals. The results of these audits are
recorded in our Radiation Protection File. If the quality targets set out
above are not met, she is responsible for taking action to ensure that overall quality is improved.
clinical evaluation of x-rays
Clinical evaluation of x-ray images is carried out by
Joanne Giddy (who is also the RPS and is the only referrer/operator/practitioner
for the purposes of IRMER at this practice) immediately after the image is
taken. This is possible since the image is captured using a digital CCD
system and it is therefore available to view only a few seconds after the
exposure. Any caries revealed or other findings relevant to the patient's
management or prognosis are recorded in the patient's electronic notes during
the course of or immediately after their appointment.
referral and justification of x-rays
This practice follows the procedures outlined in ss2.29
to 2.35 of "Guidance notes for dental practitioners on the safe use of x-ray
equipment" published by the National Radiological Protection Board in 2001 and
the more detailed good practice guidelines entitled "Selection Criteria for
Dental Radiography" published by the Faculty of General Dental Practitioners
(UK) of the Royal College of Surgeons of England, 2nd edition (2004).
This practice does not carry out x-ray exposures for purely medico-legal reasons nor
at the request of third parties.
risk assessment
The practice is required to carry out, record and periodically review a risk assessment
for work with x-rays.
It is the responsibility of the RPS (Joanne Giddy) to ensure that the risk assessment is reviewed, updated
and any necessary action taken to maintain safe working practices and/or compliance with current
regulations. The risk assessment must be reviewed:
a) whenever new radiographic techniques or equipment are introduced for the first time (eg first
use of panoramic radiography or cephalometry);
b) whenever there are changes to the process or methods of work;
c) whenever new legislation is introduced relating to radiography.
The latest risk assessment is published on this website.
training
All new members of the team receive essential training in relation to the local rules as
part of their induction process. The RPS (Joanne Giddy) must ensure that existing members of the team maintain their knowledge of the local
rules and that appropriate training is given either by way of providing written or oral instructions, or at
a special training session.
disposal of radiographic waste
Since the practice now uses a digital x-ray system, we no longer produce the associated
waste products: lead foil from x-ray wrappers, x-ray developer solution and x-ray fixer solution.
top
back to "about" menu
|